Main body
|
High-strength engineering plastic rack, light,
beautiful and corrosion resistant
|
Scope of application
|
adult, child
|
Gas source
|
O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
|
Flow meter
|
O2 : 0.1~1.0L/min 1.1~10L/min
|
N20: 0.1~1.0L/min 1.1~10L/min
|
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
|
Flow rate of rapid oxygen supply
|
25~75L/min
|
Low oxygen pressure alarm (Oxygen Alarm)
|
There will be sound alarm when the oxygen
pressure < 0.2MPa
|
Vaporizer (Vaporizer Definition and Vaporizer Construction)
|
it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
|
Respiratory circuit
|
working mode: all-close, semi-close, semi-open
APL: 0.5~7 kPa
|
Respiratory Bellows
|
bellows for adults, tidal volume range:0~1500 ml
bellows for children, tidal volume range:0~300 ml
anyone can be chosen for application as
required by the customer
|
Ventilator (Ventilator Definition and Ventilator Function)
|
Display mode
|
high-definition 5.7” LCD display
|
Ventilation mode
|
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
|
Ventilation parameters
|
Tidal volume
|
20~1500 ml
|
Rate
|
1~99 bpm
|
SIMV rate
|
2~20 bpm
|
I:E
|
4:1~1:8
|
Inspiratory trigger pressure
|
-1.0~2.0 kPa
|
PEEP
|
0~2.0 kPa
|
Pressure range
|
1.0~6.0 kPa
|
SIGH
|
1.5 times the inspiratory time
(70-150 adjustable)
|
Parameters for ventilation monitoring
|
tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform,PEEP,inspiratory trigger pressure, Inspiratory platform, autonomous inspiratory rate, lung compliance, airway resistance, flow-volume loop, pressure-volume loop
|
Monitoring of oxygen concentration
|
(21%~100%)
|
Security Alarm System
|
Oxygen Concentration alarm
|
upper limit setting range 21%~100%
low limit setting range 10%~80%
|
Airway pressure alarm
|
upper limit setting range 0.3~6.0 kPa
low limit setting range 0.2~5.0 kPa
|
Per-minute ventilation volume alarm
|
upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
|
Sustained high-pressure alarm
|
it will give alarm when stress have consistently been higher than 2.5 kPa
|
Suffocation alarm
|
it will give sound and light alarm if there is no tidal volume input for 15 seconds
|
Power alarm
|
Power
|
AC 220V±10% 50Hz±1Hz(UPS, Storage battery)
|
(Mechanical Arms)
|
(End-expiratory CO2 monitoring)
Items within brackets are optional.
|