Best 5-part hematology analyzers Mindray BC 5000 - Medsinglong
BC-5000
Auto Hematology Analyzer
Technical Specifications
Principles
Impedance method for RBC and PLT counting
Cyanide free reagent for hemoglobin test
Flow Cytometry (FCM) + Tri-angle laser scatter + Chemical dye
Parameters
23 parameters: WBC, Lym%, Mon%, Neu%, Bas%, Eos%, Lym#,
Mon#, Neu#, Eos#, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC,
RDW-CV, RDW-SD, PLT, MPV, PDW, PCT
3 histograms for WBC, RBC and PLT
Reagent
M-52D Diluent, M-52DIFF Lyse, M-52LH Lyse, Probe Cleanser
Performance
Parameter Linearity Range Precision Carryover
WBC 0-100×10 9 /L ≤2% (4-15×10 9 /L) ≤0.5%
RBC 0-8×1012/L ≤1.5% (3.5-6.0×1012/L) ≤0.5%
HGB 0-250g/L ≤1.5% (110-180g/L) ≤0.6%
PLT 0-1000×10 9 /L ≤4.0% (100-500×10 9 /L) ≤1.0%
Sample Volume
Prediluted mode 20 μL
Whole blood mode 15 μL
Capillary whole blood mode 15 μL
Throughput
40 samples per hour
Display
10.4 inch TFT Touch Screen
MSL TEAM picture